ECT-F electronic controller
The interactive display may be customized by hospital, pharmaceutical and scientific professionals and enables an immediate check of all main settings for a fast trouble shooting of eventual problems (Doctor View).
Special functions such as Stop-on-door and Advanced defrosting functions allow an efficient managing of the defrost cycles, reducing thermal variations and consumption.
A list of the last 16 alarms is available from the screen, with signals for: - Open door - High temperature - Low temperature - Power failure (signaled only when power is restored and if temperature has exceeded the pre-set limits) - Clogged condenser
The password identifies two separate access level to the controller (end user and Service), thus protecting the working parameters, the On/Off main switch (on request) and the operating set-point of the unit.
Standard special functions for power saving
Advanced Defrost Monitoring: thanks to the continuous monitoring of evaporator maximum and minimum temperatures during every working cycle, in case a too high ice thickness level is detected on the evaporator, the controller will activate a defrost.
Advantages: lower power consumption and greater guarantee of temperature uniformity.
Stop-on-door: this function enables to avoid an excessive decrease of the evaporator temperature due to switching off of the fan during door opening. It affects the compressor turn off in case of a too high evaporator overcooling.
Advantages: minor thermal shock, after door opening
ECT-F PLUS microprocessor controller
Electronic controller, with 0,1 °C resolution, fitted with backlit alphanumeric LCD display with a 5 languages menu.
Clear and precise information concerning the working status of the refrigerator is displayed, and so are all the detected critical temperatures
100% interactivity
A simple, protected and differentiated surfing guarantees access to necessary and detailed information, both for end users and for service.
In addition to ECT-F controller features,
the ECT-F PLUS electronic control board includes:
- Back-up batteries for power failure alarm up to 48H (24H for 2 temperatures models)
- High temperature alarm due to power failure
- Temperature display during power failure
- Min/Max temperature recording function
- Potential free contact for remote alarms
ECT-F TOUCH microprocessor controller
Supervision electronics at the highest level ensures high security for specialized medical and scientific refrigerators with a simple and intuitive interface.
The ECT-F TOUCH range refrigerators and freezers include as standard the recording of temperatures and alarms related to the latest 10 years; these data can be downloaded by means of the supplied SD card and software, or directly with an USB pen from the front of the equipment.
A simple, protected and differentiated surfing guarantees access to necessary and detailed information, both for end users and for service.
- TFT 7’’ touch screen display easily readable from the distance. The home page shows main information while a real-time customizable graphic displays temperatures
- Data recording on SD Card and in a flash memory sampling every 30 sec.
- USB port for software upgrade and data downloading
- Ethernet RJ485 port for connection to LAN net
- SIM Card slot for GSM module
- Dry contact to remote alarms
- DMLP Touch Digital Monitor for a continuous recording by means of an independent PT100 probe
- Mod-bus communication protocol for the connection to independent pre-existent centralized systems
- WLS integrated module: the assigning of a unique IP address for any equipment
100% SAFETY
Complete control of the refrigerator and recording of all event during the life cycle of the medical refrigerator.
- Double feeding, double compressor relays and double defrost: this will guarantee the working of the equipment even in case of irreversible failure
- Alarm and failure reporting
- Protection against high condenser temperature
- Protection against low evaporator temperature
- Product protection against thermoregulation probe failure
- The product will remain safe even in case of faulty conditions: protection against freezing or abnormal temperature rise
- Selective multi-user management for functions
100% DIAGNOSTIC
Self-diagnostics, functional tests and failures prevention